Cineradiographic evaluation of Björk-Shiley monostrut prosthetic valve
スポンサーリンク
概要
- 論文の詳細を見る
Functional property of the Björk-Shiley monostrut prosthetic valve implanted in the mitral position of 30 patients was investigated by cineradiography. The maximum opening angle of the tilting disc was 70.6±4.6° There was no change in the maximum opening angle, opening time, closing time and tilting disk angle of the disc observed during follow up period. It is noteworthy that the rotation of the valve leaflet is not observed in long term.
- 一般社団法人 日本人工臓器学会の論文
著者
-
山口 寛
Second Department of Surgery, Tokyo Medical College
-
矢尾 善英
Second Department of Surgery, Tokyo Medical College
-
長田 一仁
Second Department of Surgery, Tokyo Medical College
-
藤川 正
Second Department of Surgery, Tokyo Medical College
-
平山 哲三
Second Department of Surgery, Tokyo Medical College
-
石川 幹夫
Second Department of Surgery, Tokyo Medical College
-
北村 昌之
Second Department of Surgery, Tokyo Medical College
-
石丸 新
Second Department of Surgery, Tokyo Medical College
-
古川 欽一
Second Department of Surgery, Tokyo Medical College