Clinical evaluation of the new PES hemoconcentrator LH-420 P
スポンサーリンク
概要
- 論文の詳細を見る
A new hemoconcentrator (LH-420 P) was employed to control plasma volume in 30 patients during cardiopulmonary bypass operation. LH-420 P was made of polyethersulfone membrane, and had a smaller priming volume than the current model (Hemocon®). At 350-400 torr and a flow rate of 250-300 ml/min, the ultrafiltration rate was 82.4±27.3 ml/min, which was 1.8 times higher than Hemocon®. The permeable properties of the blood components were equal to those of the previous device, and blood cells and plasma proteins were concentrated with ease. Thus, LH-420 P was useful for plasma water removal by ultrafiltration in association with cardiopulmonary bypass.
- 一般社団法人 日本人工臓器学会の論文
著者
-
石原 義久
Central Res Inst of Japan Medical Supply Co., Ltd.
-
伴 敏彦
Dept of Cardiovasc Surg, Kyoto University Medical School
-
野本 慎一
Dept of Cardiovasc Surg, Kyoto University Medical School
-
岡本 好史
Dept of Cardiovasc Surg, Kyoto University Medical School
-
吉本 文雄
Central Res Inst of Japan Medical Supply Co., Ltd.
-
土谷 太郎
Central Res Inst of Japan Medical Supply Co., Ltd.