- 論文の詳細を見る
To evaluate the late results of valve replacement with Hall-Kaster (HK), Omniscience (OS), Bicer-Val (BV) and Björk-Shiley C-C (CC) valve, which we used for a short period only, we compared actuarial survival (AS), reoperation free (RF), and event free (EF) rates of these prostheses to those of the St. Jude Medical (SJM), valve prosthesis. Number of patients: HK20, OS68, BV37, CC28, SJM1284. AS: HK47.4%*, OS76.8%*, BV66.7%*, CC77.9%, SJM84.4%, RF: HK92.9%, OS89.7%*, BV79.9%0*, CC91.6%, SJM98.7%, EF: HK46.6%*, OS66.8%*, BV58.1 %*, CC68.7%*, SJM82.7% (* p<0.05 vs SJM). These results justify our decision to stop the usage of HK, OS, BV and CC.
- 一般社団法人 日本人工臓器学会の論文
著者
-
八巻 文貴
Depatment of Cardiovascular Surgery, The Heart Institute of Japan Tokyo Women's Medical College
-
中野 清治
Depatment of Cardiovascular Surgery, The Heart Institute of Japan Tokyo Women's Medical College
-
遠藤 真弘
Depatment of Cardiovascular Surgery, The Heart Institute of Japan Tokyo Women's Medical College
-
橋本 明政
Depatment of Cardiovascular Surgery, The Heart Institute of Japan Tokyo Women's Medical College
-
小柳 仁
Depatment of Cardiovascular Surgery, The Heart Institute of Japan Tokyo Women's Medical College
-
小柳 仁
Depatment of Cardiovascular Surgery, The Heart Institute of Japan Tokyo Women's Medical College