A New Method for Plotting Exponentially Scanned Spectraon a Linear Mass Scale
スポンサーリンク
概要
- 論文の詳細を見る
A new method for plotting linearized spectra from an exponentially scanned magnetic mass spectrometer on a strip chart recorder has been developed. The chart paper is advanced by a pulse motor which is fed from V/F converter, the input signal of which is proportional to the square of the scanning signal. The scanning speed can be varied without changing the linear mass scale on chart. The linearity is sufficient to interpolate mass numbers over a span of one hundred mass units.
- 日本質量分析学会の論文
著者
-
TAKESHITA ISAO
Central Research Laboratory Matsushita Electric Industrial Co.,Ltd.
-
TAKESHITA ISAO
Central Research Laboratory, Matsushita Electric Industrial Co., Ltd Yakumonakamachi,
-
TANAKA HIDEO
Central Research Laboratory, Matsushita Electric Industrial Co., Ltd Yakumonakamachi,
-
Shino Taichi