スパークイオン源形固体質量分析器による金属中の不純物定量分析法の検討
スポンサーリンク
概要
- 論文の詳細を見る
Attempts have been made to improve the precision and accuracy of the analysis of metals performed with spark source mass spectrograph by using some standard samples(aluminum, steel, copper and lead matrix)that contained impurities in concentrations ranging from a few thousan down to ten ppm. Reproducibility expressed in terms of coefficient of variation of the analytical results obtained on repeated exposures can attain 5 to 15 per cent(Tables 1 through 4)provided that all source oonditions are maintained constant, especially the spark gap at a rather large value of ca. 50μ and all plots for calculation curves are corrected for line width variation. From the analytical results thus obtained and further subjected to correction for mass dependence of emulsion sensitivity, the relative sensitivity factors of impurity elements have been. determined at the values ranging from 0.48 to 3.6(Fe=1)(Table6. cf. Table5). The experimental relative sensitivity factors agree within ± 30% with the values calulated by the relation proposed by Kai et al. in which the sensitivity factor of an element is inversely proportional to the product of its ionization potential and sublimation energy(Fig.4).
- 日本質量分析学会の論文
著者
関連論文
- スパークイオン源固体質量分析における乾板によるイオン量測定精度の検討
- イオン衝撃形固体質量分析器における二次イオン特性
- 溶液の定量分光分析における回転炭素電極の改良 パラフィン処理による
- スパークイオン源形固体質量分析器による金属中の不純物定量分析法の検討
- 電子計算機による固体質量スペクトルの会話形デ-タ解析
- 質量分析用Ilford Q2乾板の乳剤感度の質量依存性