Microcomputer Scanning System for Respiratory Mass Spectrometer
スポンサーリンク
概要
- 論文の詳細を見る
The function of the sytem is to determine step voltages for scanning gases to bemeasured, and to make simultaneous and continuous measurement of the specified mass ions by a rapid and repeated supply of the step voltages and gate pulses. The system is composed of a microcomputer scanning system and an analogue gate assembly. The microcomputer system is composed of an A/D converter(ADC550-10-E), a D/A converter(DAC371V-8)and a SDK-80system(8080CPU, ROM2KB, RAM1KB and a programmable peripheral interface 8255). The 8255 plays an important role as an I/O port in the construction of the interface. The design construction of the system is illustrated and the operation of the system is described in details. Some technical points are examined and discussed on how to specify the scanning step voltages, and on how to adjust or maintain the optimum step voltages.
- 日本質量分析学会の論文
著者
-
Nishi Isao
Faculty of Science and Technology, Science University of Tokyo
-
Tomizawa Giichi
Faculty of Science and Technology, Science University of Tokyo
-
Tatsuta Lumi
Faculty of Economics, The Dokkyo University
関連論文
- Microcomputer Scanning System for Respiratory Mass Spectrometer
- An Automatic Correction Circuit for Respiratory Mass Spectrometers
- A Mass Spectrometric System Equipped with Multi-Collectors and Peak Selectors for Continuous Gas Analysis