A simplified apparatus has been constructed for determining the purity of organic compounds by time-temperature freezing or melting curves. The sample cell is small and the amount of the sample needed in only 0.2 m<I>l</I> or 2 m<I>l</I>. The thermometers used are a Shimadzu platinum resistance thermometer calibrated by the measurement of fixed points and a chromel <I>p</I>-constantan thermocouple calibrated against a standard platinum thermometer with a precision better than ±0.03°. Purities of several hydrocarbons were determined by measuring their freezing and melting curves.