Strain Gauge Electromicrobalance for Surface Tension Measurement
スポンサーリンク
概要
- 論文の詳細を見る
Two types of apparatus using an unbonded strain gauge of a cantilever type for the measurement of surface tension were constructed. One is a sensitive electromicrobalance equipped with an extended arm attached to the cantilever of the strain gauge. The apparatus proved suitable as a surface balance in a completely closed system. By this apparatus, the measurement of surface tension ranging from 0 to 74 dyn/cm is possible with the accuracy of ±0.10 dyn/cm. The other type was constructed by connecting the sensing lever of strain gauge to the arm of a torsion balance with a strip of phosphor bronze. This device together with a counterbalance on the opposite side of the arm also proved to be useful as a tensiometer. Accuracy of this apparatus was a little better than that of the extended arm type balance, although a larger space is required. Both apparatus are inexpensive, easy to construct and simple to operate, and are suitable for the self-recording of time dependence of surface tension or weight change in general with sufficient accuracy. Equations proposed for the time dependence of surface tension of surfactant solution were checked by using the strain gauge tensiometer of the extended arm type.