Survey of dose distribution in patients due to operation of CT machines.
スポンサーリンク
概要
- 論文の詳細を見る
This paper deals with dose measurement in gonad and eyes with self-fabricated thermoluminescent dosimeters during CT scanning for brain. A female Rando phantom was used as the reference. The experimental results obtained from several patients show that for brain scan, the average eye dose was 169.5±204.1mrad/slice and gonad dose 0.20±0.15mrad/slice for male, and that for female was 195.3±193.2mrad/slice for eyes and 0.04±0.03mrad/slice for gonad. The results are also compared with that obtained from other models of CT scanner.
- 日本保健物理学会の論文
著者
-
WENG PAO-SHAN
Institute of Nuclear Science National Tsing Hua University
-
SU Ling-Nah
Department of Nuclear Medicine; Kaohsiung Medical College
-
HSU Pin-Chieh
Institute of Nuclear Science, National Tsing Hua University
関連論文
- Natural Radiation Background in Metropolitan Taipei
- Radon Concentration in Two Largest Cities in Semitropical Taiwan
- X-radiation from television receivers and video display terminals.
- Personal exposure rates as a function of reactor power.