Case studies on the effect of exercise and hot water submersion on intracardiac temperature and the performance of a pacemaker which varies pacing rate based on temperature.
スポンサーリンク
概要
- 論文の詳細を見る
The effectiveness of using blood temperature change as an indicator to automatically vary heart rate physiologically was evaluated in 3 patients implanted with Model Sensor Kelvin 500 (Cook Pacemaker Corporation, Leechburg, PA, USA) pacemakers. Each patient performed two block-randomized treadmill exercise tests: one while programmed for temperature-based, rate-modulated pacing and the other while programmed without rate modulation. In 1 pacemaker patient and 4 volunteers, heart rates were recorded during exposure to a hot water bath.Blood temperature measured at 10sec intervals and pacing rate measured at 1min intervals were telemetered to a diagnostic programmer and data collector for storage and transfer to a computer. Observation comments and EGG-derived heart rates were manually recorded.The temperature-based pacemaker was shown to respond promptly not only to physical exertion but also to emotionally caused stress and submersion in a hot bath. These events cause increased heart rate in the normal heart. Using a suitable algorithm to process the measurement of blood temperature, it was possible to produce appropriate pacing rates in paced patients.
- International Heart Journal刊行会の論文
著者
-
Fujita Tamotsu
Okamura Memorial Hospital
-
Okamura Hiroshi
Okamura Memorial Hospital
-
Kitoh Osamu
Okamura Memorial Hospital
-
SMITH Heidi
Hillenbrand Biomedical Engineering Center, Purdue University
-
FEARNOT Neal
Hillenbrand Biomedical Engineering Center, Purdue University
-
CALDERINI Mark
Cook Pacemaker Corp.
関連論文
- An Early Predictor of Left Ventricular Remodeling After Reperfused Anterior Acute Myocardial Infarction : Peak E-wave Velocity / Flow Propagation Velocity
- -0040-PROGRESSIVE DOWN REGULATION OF BETAADRENOCEPTOR (AR) IN HUMAN MYOCARDIUM WITH ADVANCEMENT OF HEART FAILURE
- Case studies on the effect of exercise and hot water submersion on intracardiac temperature and the performance of a pacemaker which varies pacing rate based on temperature.