Comparison of a Semi-Automated Sphygmomanometer with the Clinical Sphygmomanometer
スポンサーリンク
概要
- 論文の詳細を見る
A semi-automated sphygmomanometer Stanley B-200 has been compared with the clinical sphygmomanometer in 500 college boys by using a blind simultaneous method. Stanley B-200 has shown significantly higher systolic blood pressure (14.8mmHg, P<0.05) and diastolic blood pressure (3.8mmHg, P<0.05). This device also recorded 15 false positive systolic and 12 false negative diastolic results (P<0.05).
- International Heart Journal刊行会の論文
著者
-
ILYAS Muhammad
Cardiology Division, Muhammadi Hospital
-
PERACHA Masood
Cardiology Division, Muhammadi Hospital
-
RAHMAN Inayatur
Cardiology Division, Muhammadi Hospital