An Explanation of the Discrepancies in Pattern Coefficients Obtained with Different Types of Mass Spectrometers
スポンサーリンク
概要
- 論文の詳細を見る
The author proposes a new approach to deriving the pattern coefficients of mass spectrometers in terms of such presumably measurable quantities as the partial ionization cross section and others. Firstly, the Intrinsic Pattern Coefficients that are inherent to an ion source are derived in terms of (l vs V) distribution of ionizing electrons and the partial ionization cross section of a sample gas. The (l vs V) distribution gives the transit path length of ionizing electrons as a function of the electron energy. Secondly, the amount of each fragment ion which is lost in the mass analyzing region is estimated. This approach was applied to three vacuum analyzers. The mutual differences in pattern coefficients that exist among them were successfully accounted for.
- 日本質量分析学会の論文
著者
-
Mima Hiroshi
Faculty Of Engineering Osaka City Univercity
-
MIMA HIROSHI
Faculty of Engineering, Osaka City University
関連論文
- Paramagnetic Resonance Linewidth of Cr^ in [Co en_3]Cl_3・nH_2O^*
- An Explanation of the Discrepancies in Pattern Coefficients Obtained with Different Types of Mass Spectrometers