Oxygenation to Bovine Blood in Artificial Heart and Lung Using Vibrating Flow Pump: Experiment and Numerical Analysis Based on Non-Newtonian Model
スポンサーリンク
概要
- 論文の詳細を見る
In this study, we construct an experimental apparatus for a prototype artificial heart and lung (AHL) by installing hollow fibers into the cylindrical tube of the vibrating flow pump (VFP). The oxygenation characteristics are investigated both by experiments using bovine blood and by numerical analyses based on the computational fluid dynamics. The analyses are carried out at the Reynolds numbers Re ranged from O(1) to O(103), which are determined based on the experimental conditions. The blood flow and the diffusion of oxygen gas are analyzed based on the Newtonian/non-Newtonian, unsteady, incompressible and axisymmetric Navier-Stokes equations, and the advection-diffusion equation. The results show that the oxygenation rate increases in proportion to Re1/3, where the phenomenon corresponds to the decreasing thickness of the concentration boundary layer with Re. Although the effects of the vibrating flow and the rheology of the blood are clearly appeared on the velocity field, their effects on the gas exchange are relatively small at the ranges of prescribed Reynolds numbers. Furthermore, the numerical results in terms of the oxygenation rate are compared with the experimental ones. The basic design data of VFP were accumulated for the development of AHL in the clinical applications.
論文 | ランダム
- Improvement of a Capillary Electrophoresis-Chemiluminescence Detection System for Using a Polyacrylamide-Coated Capillary
- Application of Microchip Capillary Electrophoresis with Chemiluminescence Detection to an Analysis for Transition-Metal Ions
- キャピラリー電気泳動-化学発光検出法の免疫分析への応用 : 色素包括リポソーム標識剤の化学発光検出
- 10-フェナントロリン-過酸化水素-オスミウム(VIII)化学初光を用いるα-アミノ酸の定量
- Immunoassay Using Chemiluminescence Detection of Dyestuff-Containing Liposomes as a Labeling Reagent