Determination of Citrate Using Immobilized Citrate Lyase and Malate Dehydrogenase in a Flow System and Its Application to Analyze Citrate Content of Beverages
スポンサーリンク
概要
- 論文の詳細を見る
The quantity of citrate was determined using apparatus comprised of a reactor with immobilized citrate lyase and malate dehydrogenase in the flow line. The decrease of, β-nicotinamide adenine dinucleotide, reduced form by enzymatic reactions was spectrophotometrically detected as a negative peak. The maximum peak area was obtained at pH 7.6 when the pH of the carrier consisting of phosphate buffer ranged from 6.6 to 8.0. Various buffer types were also examined as carrier media at pH 7.6 and phosphate buffer showed the maximum peak area. When the carrier composed of phosphate buffer (0.1M, pH 7.6) was used, the calibration curve for citrate was linear in the range 0.5-100μM (r=1.000). The detection limit (S/N=3) was 0.2μM. The relative standard deviation of the pear area at 10μM was 2.2% (n=7). This method was applied to analyze citrate in beverages, and citrate content determined by this method agreed with that determined by a commercially available test kit.
- 公益社団法人日本薬学会の論文
- 2009-02-01
著者
-
Mori Hisakazu
Faculty Of Pharmacy Keio University
-
Kadowaki Takashi
Faculty of Pharmacy, Keio University
-
Kadowaki Takashi
Faculty Of Pharmacy Keio University