Polarization-Tunable X-Ray Optics for Synchrotron Radiation
スポンサーリンク
概要
- 論文の詳細を見る
A new type of polarization-tunable X-ray optics for synchrotron radiation is proposed. The optics consists of a double-crystal monochromator and a transmission-type phase plate, The monochromator and the phase plate are arranged in a (+, -) parallel setting, and their scattering plane is inclined by α°(α ≠ 0°, 90°) from the horizontal plane. The optics were tested on beamline BL-15C of the Photon Factory; circular and vertical polarizations were successfully produced at a wavelength of 0.1026 nm. The polarization transformation efficiency at the phase plate was as high as 70%.
- 社団法人応用物理学会の論文
- 1997-05-15
著者
-
Hirano Keiichi
Photon Factory Institute Of Materials Structure Science High Energy Acccelerator Research Organizati
-
Hirano Keiichi
Photon Factory National Laboratory For High Energy Physics
関連論文
- Determination of the Density and the Thickness of SiO_2 Films Using Extremely Asymmetric X-Ray Diffraction
- Determination of the Stokes-Poincare Parameters for a Synchrotron X-Ray Beam by Multiple Bragg Scattering
- Polarization-Tunable X-Ray Optics for Synchrotron Radiation