Measurements of Pressure of Extracorporeal Shock Wave Lithotripter Using Pressure-Sensitive Papers
スポンサーリンク
概要
- 論文の詳細を見る
This paper describes measurements of pressures at the focal region of the extracorporeal shock wave lithotripter (ESWL) using pressure-sensitive papers. At the focal region of ESWL, ordinary hydrophones are quickly damaged, because of very high pressures. Recently, measurements of pressure at the focal region of ESWL using pressure-sensitive paper have been advised. Therefore, we have studied the effectiveness of pressure-sensitive papers in the measurement of high acoustic pressures at the focal region of ESWL.
- 社団法人応用物理学会の論文
- 1993-05-30
著者
-
IDE Masao
Faculty of Engineering, Musashi Institute of Technology
-
Ide Masao
Faculty Of Engineering Musashi Institute Of Technology
-
INOSE Naoto
Faculty of Engineering, Musashi Institute of Technology
-
Inose Naoto
Faculty Of Engineering Musashi Institute Of Technology
関連論文
- Effects of Ultrasound on Extraction of Saponin from Ginseng
- Effect of Ultrasonic Intensity on Sprouting
- Measurement of Electric Field Dsitributions in Solid Dielectrics Using Ultrasonic Waves : Acoustical Measurements and Instrumentation
- Variation of Pressure Waveforms in Measurements of Extracorporeal Shock Wave Lithotripter
- A Miniature Hydrophone for High Acoustic Pressures : Ultrasonic Motor, Actuator and Transducer
- Measurements of Pressure of Extracorporeal Shock Wave Lithotripter Using Pressure-Sensitive Papers