Pressure Distribution for Piezoelectric Extracorporeal Shock Wave Lithotripsy
スポンサーリンク
概要
- 論文の詳細を見る
The objective of this study is to develop a safer and more effective extracorporeal shock wave lithotripter. The first stage of the study shows the sound pressure field of the shock wave made by an ECHOLITH ESL-500A. The sound pressure distribution is in a ring configuration on a 60 mm plane in front of the focal plane. As the plane approaches the focal plane, the sound pressure relatively increases at the cross point with the axis of the transducer and decreases at the ring. The focal zone is 2.5 mm × 16.1 mm at 60 V driving voltage and 1.8 mm × 14.2 mm at "INTENSITY 2." In the next stage we propose a method for changing the field by electronic driving control of each piezoceramic element for effective therapy. The focal zone can be changed from 3.1 mm × 19.1 mm to 3.9 mm × 32.4 mm at 60 V driving voltage with this method. These focal zones are calculated by means of computer simulation.
- 社団法人応用物理学会の論文
- 1993-05-30
著者
-
Iwama Nobuyuki
Medical Engineering Laboratory Toshiba Corporation
-
Iwama Nobuyuki
Medical Engineering Laboratory Toshiba Co.
-
YANAGIDA Yuji
Medical Engineering Laboratory, Toshiba Corporation
-
Yanagida Yuji
Medical Engineering Laboratory Toshiba Corporation
-
Okazaki Kiyoshi
Medical Engineering Laboratory Toshiba Corporation
-
Okazaki Kiyoshi
Medical Engineering Laboratory Toshiba Co.
関連論文
- Fundamental Study in Extracorporeal Shock Wave Lithotripsy Using Piezoceramics : High Power Ultrasonics
- Measurement of Shock Wave Pressure Distribution of Extracorporeal Shock Wave Lithotripter Using Pressure-Sensitive Paper : Practical Usefulness and Limitations
- Effect of Controlled Sound Pressure Field on Fragmentation in Piezoelectric Extracorporeal Shock Wave Lithotripsy
- Pressure Distribution for Piezoelectric Extracorporeal Shock Wave Lithotripsy
- Process of Change from Ultrasonic Wave to Shock Wave in Extracorporeal Shock Wave Lithotripsy Using Piezoceramic Transducer : High Power Ultrasonics