A simple volumetric apparatus for measuring the rate of gas exchange in microorganisms
スポンサーリンク
概要
- 論文の詳細を見る
A new simple volumetric apparatus for measuring the respiration, fermentation and photosynthesis rates in microorganisms was constructed. The apparatus, its operation and some data obtained with it are presented. This apparatus 1) has an open-capillary constant-pressure volumeter without a compensation chamber, 2) has a magnetic stirrer for stirring the reaction mixture, 3) can be used to measure with a sensitivity of 0.85 μl volume change per one mm reading without a microscope, 4) simultaneously uses seven volumeters, one of which is set as a thermobarometer, 5) is completely immersed in a water bath to minimize the effect of ambient temperature, and 6) simplifies calculation.
- 日本植物生理学会の論文
著者
関連論文
- A simple volumetric apparatus for measuring the rate of gas exchange in microorganisms
- Accumulation of Oxygen Gas in Excised Maize Root on Aging in Water, and Water Uptake by the Root