エチルパラチオン, マラソンのガスクロマイトグラフィーによる定量 : 特にガスクロマトグラフィー諸条件がそれらの熱分解におよぼす影響について
スポンサーリンク
概要
- 論文の詳細を見る
Thermal degradation brought with gas chromatographic analysis of Ethyl parathion (EP) and Malathion (M) are quantitatively studied. Under the various temperature of injection part and column, and various retention time, the factors which should effect on the degradation of the samples are studied. The stability of the samples are determined by comparing with stable internal standard substance such as di-n-butyl phthalate (BP). Calibration curves for the assay are prepared with a column containing 1.5% neopentyl glycol succinate (NGS) as the stationary phase.
- 公益社団法人日本薬学会の論文
- 1967-10-30