空中およびPhantom中における高Energy電子線線量の比較
スポンサーリンク
概要
- 論文の詳細を見る
Electron doses in air and phantoms were measured using an ionization chamber, JAPM substandard dosimeter. The ratios of chamber readings in air to those in Mix-Dp pantom decreased gradually with decrease in electron energy. For 4 to 10 Mev electrons, dose readings in air were 10 to 15 percent lower than those in Mix-Dp phantom. These differences exceeded the permissible error of dose determination in radiation therapy. Dose readings in Lucite phantom were abut 3 percent higher than those in Mix-Dp phantom regardless of electron energy. These results lindicated that Lucite phantom can be used for relative dosimetry such as depth dose determination, but absolute doses should be measured in tissue equivalent phantom.
- 公益社団法人日本放射線技術学会の論文
- 1974-01-31