X線管焦点寸法の管電流による変化
スポンサーリンク
概要
- 論文の詳細を見る
At present, the dimensions of focal spot is determined by means of X-ray pin-hole cameras and resoluti on method. But this method is not practical for routine work. The magnified radiographic image of star test pattern is a useful determination of effective focal spot size. We investigated the change of effective focal spot size by means of the star test pattern method. In the result of exeriment, the effective focal spot size is enlarged by increasing milliamperage. But this phenomenon has been expected. The rate of enlargement is different from each X-ray tube. Ttis phenomenon is caused by convergence effect of thermoelectron. The effective focal spot size is present the minimum value by milliamperage. When the radiographic image is maked, the milliamperage value must be chosen a minimum effective focal spot size.