増感紙-フイルム系における濃度のMTF
スポンサーリンク
概要
- 論文の詳細を見る
Screen-film system is generally used, in radiographic system. Sharpness of screen-film system is frequently measured by means of Modulation Transfer Function (MTF). But it is prospected for the reason of irradiation, that MTF is changed by film density. Therefore, we evaluted the relation of MTF with film density concerning screen-film system. And we intended to know the optimum density of x-ray picture. In the result of experiment, the higher film density, the poorer MTF is. Consequently, X-ray film MTF in low density is better than in high density. It was found that MTF is changed by film density. Therefore, MTF exist as function of density. Many line density function is developed by a line spread function. The study of the irradiation is a task affter this.
- 公益社団法人日本放射線技術学会の論文
- 1972-01-20
著者
-
永田 耕作
宮崎県立宮崎病院放射線科
-
河野 誠一
宮崎県立宮崎病院放射線科
-
岩切 国雄
県立宮崎病院放射線科
-
稲津 博
県立宮崎病院放射線科
-
永田 耕作
県立宮崎病院放射線科
-
河野 誠一
県立宮崎病院放射線科
-
篠崎 悌五
県立宮崎病院放射線科
-
篠崎 悌五
宮崎県立宮崎病院
-
篠崎 悌五
県立宮崎病院
-
河野 誠一
県立宮崎病院
-
永田 耕作
県立宮崎病院
-
稲津 博
県立宮崎病院
-
岩切 国雄
県立宮崎病院
関連論文
- 35.線像強度分布の新しい求め方
- フィルム等方性の自現機による乱れ
- 224.フィルム等方性の自現機による乱れ(現像処理)(第28回総会)
- 13.多重シャッターの比較(九州部会)
- 24.増感紙-フィルム系におけるMTF(◇九州部会)
- 増感紙-フイルム系における濃度のMTF
- 129.全遠隔操作方式X線テレビの圧迫部分の改造とその圧迫効果についての検討
- 11.鳥啄突起の撮影法について
- 37.断層写真像の画像評価法
- 24.散乱線含有率測定法の比較(九州部会)
- 13.グリッド密度による散乱線除去効果の変化(九州部会)
- 散乱線除去効果のMTF
- 10.散乱線除去効果のMTF(撮影理論(2))(第28回総会)
- 8.多軌道断層撮影装置におけるDulac氏法の最適軌道(九州部会(第28回))(一般研究発表)
- 22.X線管角度変化による実効焦点の大きさ(◇九州部会)
- 17.反転X線写真の臨床応用(九州部会)
- 増感紙- フィルム系の濃度のM T F
- フィルム等方性の自現機による乱れ
- 絞り装置による散乱線除去効果のMTF