X線イメージインテンシファイアのコントラスト比と低空間周波数におけるMTFの落ち込み(LFD)との関係
スポンサーリンク
概要
- 論文の詳細を見る
The relationship between the drop at low frequency of the MTF and contrast ratio of the x-ray image intensifier is evaluated. The three samples have the different contrast ratios. It is shown as the function of Pb disc diameter. That relationship is called the disc transfer function (DTF). The DTF can be expressed by two exponential parts. By this it is shown that the flare of the part of the small Pb disc diameter corresponds to the LFD, and the short range effects is especially related to the LFD. Therefore, it is able to be expressed by the simple equation of the relationship between the contrast ratio and the LFD of the XRII. The LFD evaluation which is the difficult measurement can take place the contrast ratio evaluation by several Pb disc diameter which is a simple measurement.
- 公益社団法人日本放射線技術学会の論文
- 1996-04-20
著者
関連論文
- 29. 高 MTF12" X 線イメージインテンシファイアの画像評価(機器・材料-3 I.I.)
- 29. 高 MTF12" X 線イメージインテンシファイアの画像評価(第 49 回総会学術大会一般研究発表予稿)
- 221. DF における 14 インチアドバンスト・スーパ・メタル I. I. の評価(X 線装置-6 I.I. 他)
- 221. DF における 14 インチ・アドバンスト・スーパ・メタル I. I. の評価(第 48 回総会学術大会一般研究発表予稿)
- 396. 16 吋アドバンスト・スーパ・メタル I.I. の開発(I.I.)
- 396. 16 吋アドバンスト・スーパ・メタル I. I. の開発(第 46 回総会学術大会会員研究発表予稿)
- 176. 高MTF14インチX線イメージインテンシファイアの開発
- 176.高MTF14インチX線イメージインテンシファアの開発(第24回秋季学術大会 一般研究発表予稿集)
- 235.高DQE12吋アドバンストI.I.の開発
- 152 X線イメージインテンシファイア(I.I.)の交換のためのガイドライン
- 75 X線イメージ インテンシファイアの実効的なDQEと量子ノイズに関するS/N比との関係
- 75 X線イメージインテンシファイアの実効的なDQEと量子ノイズに関するS/N比の関係(第25回秋季学術大会一般研究発表予稿集)
- 175. X線イメージインテンシファイアの量子ノイズに関するS/N比とDQEとの関係
- 175.X線イメージインテンシファイアの量子ノイズに関するS/N比とDQEとの関係(第24回秋季学術大会 一般研究発表予稿集)
- X線イメージインテンシファイアのコントラスト比と低空間周波数におけるMTFの落ち込み(LFD)との関係
- 71.X線イメージインテンシファイアのDQEと残光特性による量子ノイズ評価
- 入力面寸法40cmのJタイプX線イメ-ジインテンシファイア
- 71.X線イメージインテンシファイアのDQEと残光特性による量子ノイズ評価(第23回秋季学術大会一般研究発表予稿)
- 53. X線イメージインテンシファイアのコントラスト比とLFDの関係
- JタイプX線イメ-ジインテンシファイア
- 53. X 線イメージインテンシファイアのコントラスト比と LFD の関係(第 22 回秋季学術大会一般研究発表予稿)