Study of Radiation Dose and Device Quality for Diagnostic X-ray Units
スポンサーリンク
概要
- 論文の詳細を見る
Digital kVp meter, ion chamber, and x-ray films were used to study the tube voltage, beam output, and dose distribution for 29 x-ray units. Of which 14 units are equipped with single phase power supply and 15 with inverter frequency power supply. The experimental results show that the dose is dependent on tube voltage, tube current, field size except beyond 20×20cm^2, exposure distance, and heel effect. The relationship between dose and tube voltage and exposure distance as well is non-linear, while that between dose and tube current is linear. The non-uniform and non-symmetric dose distribution depends on the heel effect of the x-ray beam. The measured voltage deviating from the default value may reach-2.4% to-9.1% depending on the power supply system used. The dose delivered from the units with inverter frequency power supply is 1.64 times higher than that with single phase power supply.
- 日本保健物理学会の論文
著者
-
Weng Pao
Nuclear Science And Technology Development Center National Tsing Hua University
-
HSU Pin-Chieh
Nuclear Science and Technology Development Center
-
Chang Jung
Radiological Technology Department Yuanpei Institute Of Science And Technology:biomedical Engineerin
-
HSU Pin
Nuclear Science and Technology Development Center, National Tsing Hua University
-
CHAO Jiunn
Nuclear Science and Technology Development Center, National Tsing Hua University
-
SU Jenn
Biomedical Engineering Department, Chung Yuan Christian University
-
Su Jenn
Biomedical Engineering Department Chung Yuan Christian University
-
Hsu Pin
Nuclear Science And Technology Development Center National Tsing Hua University
-
Chao Jiunn-hsing
Nuclear Science And Technology Development Center National Tsing Hua University
-
Chao Jiunn
Nuclear Science And Technology Development Center National Tsing Hua University
-
Chang Jung-Hua
Radiological Technology Department, Yuanpei Institute of Science and Technology:Biomedical Engineering Department, Chung Yuan Christian University
関連論文
- Effect of Teflon Covers on Thermoluminescence of CaF_2:Tm
- Study of Radiation Dose and Device Quality for Diagnostic X-ray Units
- Respone of Thermoluminescence Dosimeter CaF_2 : Tm to Low-Level Proton Beams Using Rutherford Backscattering
- Experiences Gained during Twenty Field Intercomparison Studies for Self-fabricated TLD CaSO_4 : Dy
- Reassessment of Reactor Coolant and Iodine Chemistry under Accident Conditions
- Personal exposure rates as a function of reactor power.