Dual Channel IF Substitution Measurement System for Microwave Attenuation Standard(Special Issue on Microwave and Millimeter Wave Technology)
スポンサーリンク
概要
- 論文の詳細を見る
An attenuation measurement system, which employs a traceable inductive voltage divider (IVD) at 1 kHz as a reference standard and dual channel intermediate frequency (IF) substitution method, is developed as an attenuation standard in the frequency range of 5 GHz to 12 GHz. The basic properties of the system are experimentally investigated and the expanded standard uncertainty of the system is estimated to be 0.0018dB for 20 dB and 0.026dB for 80 dB attenuation.
- 社団法人電子情報通信学会の論文
- 2003-08-01
著者
-
Suzuki Kazunari
National Metrology Institute Of Japan Nmij
-
Widarta Anton
National Metrology Institute Of Japan Nmij
-
KAWAKAMI Tomoteru
AIST