Development of a High-T_c SQUID Cryo-System for the Measurement of a Remanent Mgnetic Field of Rock(Special Issue on Superconductive Devices and Systems)
スポンサーリンク
概要
- 論文の詳細を見る
A portable cryo-system using a high-T_c SQUID for the measurement of the remanant magnetic field of a rock specimen was designed and fabricated.The sensing surface of the SQUID faces upward in our system, although the system for biomagnetics faces down.The SQUID is cooled by liquid nitrogen via a sapphire heat transfer rod.The total heat transfer of the system was measured by means of a boiling-off method and was found to be 1.65W.It was demonstrated that the system can be operated for more than 17 hours without any maintenance such as filling with liquid nitrogen.The system was applied to the measurement of the remanent magnetic field distributions of rock samples cored from deep underground.We have successfully measured the distributions.
- 社団法人電子情報通信学会の論文
- 2000-01-25
著者
-
Shin Koichi
Central Research Institute Of The Electric Power Industry
-
Tanaka Saburo
The Ecological Engineering Toyohashi University Of Technology
-
SHIMIZU Ryouji
the Ecological Engineering, Toyohashi University of Technology
-
SAITO Yusuke
the Ecological Engineering, Toyohashi University of Technology
-
Shimizu R
Tohoku Univ. Sci. Noda‐shi Jpn
-
Saito Yusuke
The Ecological Engineering Toyohashi University Of Technology