Improvement in Temperature Measurement Accuracy of Q-Branch CARS Thermometry : Effects of Spectral Resolution of Detection System
スポンサーリンク
概要
- 論文の詳細を見る
The temperature measurement accuracy of N_2 Q-branch CARS (Coherent Anti-Stokes Raman Spectroscopy) below 1000K has been improved by improving the spectral resolution of the detection system. The total spectral resolution was improved from 1. 3cm^<-1> to 0. 6cm^<-1> by dispersion magnification (×3. 6) , which was effected by adding a magnifier between a monochromator and an intensified photodiode array detector. The accuracy of temperature measurement in the temperature range of 300-800K for pressures of 0. 1-3. 0MPa has been greatly improved from ±100K, in the case without the magnifier, to ±40K. The method reported in this paper has a potential advantage for practical use because of its convenience.
- 一般社団法人日本機械学会の論文
- 1993-05-15
著者
-
Asai T
Doshisha Univ. Kyoto Jpn
-
Akihama Kazuhiro
Toyota Central Research & Development Laboratories, Inc.
-
Asai Takeshi
Toyota Central Research & Development Laboratories, Inc.
-
Akihama K
Toyota Central Res. & Dev. Lab. Inc. Aichi Jpn