表面プラズモン共鳴センサーを用いる酵素反応によるコレステロールの新規測定の研究
スポンサーリンク
概要
- 論文の詳細を見る
A surface plasmon resonance (SPR) sensor was applied to a simple and speedy measurement of cholesterol. The lower detection limit was 5.6×10^<-4> mol/l for an ethanol solution of cholesterol. The measurable range for cholesterol almost equaled that of the concentration of cholesterol in serum without any pretreatment of the serum. There were differences between the data obtained by using a gold film and those by using an enzyme film, due to the presence of hydrogen peroxide produced by a reaction between the enzyme and cholesterol. The differences gave a linear plot in the cholesterol-concentration region of 2.0×10^<-3> to 2.0×10^<-2> mol/l, and reached a constant value in the region of 4.0×10^<-2> to 5.0×10^<-2> mol/l. The 3.0×10^<-3> mol/l ascorbic acid, which largely interfered with the spectroscopic measurement of cholesterol, did not interfere with this measurement of a 1.0×10^<-2> mol/l cholesterol/ethanol solution.
- 社団法人日本分析化学会の論文
- 2001-01-05
著者
関連論文
- モレキュラーインプリント膜を用いるテトラカインの表面プラズモン共鳴センサーによる測定
- pH指示薬/ポリビニルピロリドン膜の色変化に基づくにおいセンシング
- 抗原抗体反応を用いるα-フェトプロテインの表面プラズモン共鳴センサーによる測定
- 表面プラズモン共鳴センサーを用いる酵素反応によるコレステロールの新規測定の研究
- 訂正 : 抗原抗体反応を用いるα-フェトプロテインの表面プラズモン共鳴センサーによる測定