A Three Directional Ballistocardiograph
スポンサーリンク
概要
- 論文の詳細を見る
A design was made of a new simultaneous recording device for three directional ballistocardiograms on the dog's frontal chest using a direct-body system. Physical characteristics of the pick-up device was analysed by a specially de-signed detector. Frequency response showed an accurate operation of the pick-up in a frequency range from I to 350cps. Output linearity of the pick-up was satisfactory within a maximum vibration amplitude of 0.25mm at 100cps. The anesthetized dog's ballistocardiograms obtained by the present pick-up showed a marked resemblance to that obtained from a non-anesthetized normal man except a few findings. The present device may be favorably employed to pursue the composing factors of ballistocardiographic waves on an experimental basis.
- 社団法人日本循環器学会の論文
- 1966-03-20
著者
-
Nomura Yutaka
Center For Adult Diseases
-
FUJIMOTO Junzo
Center for Adult Diseases
-
TANIMOTO Toru
Center for Adult Diseases
-
MIYAGAWA Teikichi
Center for Adult Diseases
-
NOMURA Yutaka
Cardiovascular Division, Center for Adult Diseases
-
FUJIGAKI Hajime
Cardiovascular Division, Center for Adult Deseases
-
FUJIMOTO Junzo
Cardiovascular Division, Center for Adult Deseases
-
TANIMOTO Toru
Cardiovascular Division, Center for Adult Deseases
-
MIYAGAWA Teikichi
Cardiovascular Division, Center for Adult Deseases
-
MINAMOTO Suemitsu
Department of Electronics, College of Engneering, University of Osaka Prefecture
-
YONETANI Tadanori
Department of Electronics, College of Engneering, University of Osaka Prefecture
-
Yonetani Tadanori
Department Of Electronics College Of Engneering University Of Osaka Prefecture
-
Minamoto Suemitsu
Department Of Electronics College Of Engneering University Of Osaka Prefecture
-
Fujigaki H.
Center for Adult Diseases
-
Nomura Y.
Cardiovascular Division, Center for Adult Deseases
関連論文
- The Three Directional Ballistocardiogram on Experimental Coronary Stenosis and Occlusion of the Dog
- 138. Three Directional Ballistocardiograms on Experimental Coronary Stenosis and Occlusion
- Automatic Measurement of the Electrocardiogram by Digital Computer
- Studies on the Ballistocardiogram of Experimental Coronary Insufficiency : PROCEEDINGS OF THE 25TH ANNUAL MEETING JAPANESE CIRCULATION SOCIETY (Part II)
- Digitalis Intoxication and Electrolytes : Effect of Estradiol for the Prevention of Digitalis Intoxication
- Three Directional Ballistocardiogram in Mitral Stenosis : THE VALUE OF POSTERO-ANTERIOR BALLISTOCARDIOGRAM IN THE DIAGNOSIS OF MITRAL STENOSIS AND ITS SEVERITY
- Three Directional Ballistocardiogram in Pulmonary Stenosis : Exhibits
- Three Directional Ballistocardiogram in Atrial Septal Defect
- 36) Three Directional Ballistocardiogram in Atrial Septal Defect : 第18回日本循環器学会近畿地方学会総会
- A Three Directional Ballistocardiograph
- 237) Three Directional Ballistocardiogram in Mitral Valvular Dresses
- Ballistocardiographic Studies on Hypertension (II)
- 50) The Diagnostic Method for Coronary Sclerosis with Clinical Examination(Proceedings of the 24th Annual Meeting, The Japanese Circulation Society [Part I])