<原著>自動視野計Octopusの中心視野測定における視標サイズ, 背景輝度の影響について
スポンサーリンク
概要
- 論文の詳細を見る
The physiological character of the pericentral visual field near the fixation point is not so well understood as in the central 30° visual field. In this paper a report is made on the influences of stimulus size and background illumination on the sensitivity of this area studied in normal subjects and patients with several eye diseases. According to the SARGON program of the automated perimeter Octopus 201,49 test points on a 2° grid in the pericentral area were tested in normal subjects. These test points were examined under the conditions of stimulus sizes 1,3 and 5 (the visual angles of 0.108°, 0.431° and 1.724°) and under the background luminances of 0.4,4 and 31.5 asb. In the normal subjects the sensitivity curve of this pericentral area was rather flat when large stimulus sizes were used. But within the range from 0.4 to 31.5 asb the background illumination did not affect the sensitivity curve of this area so remarkably as the stimulus size. Differential light sensitivity decreased with age and its decrease was more evident under the condition of smaller stimulus sizes. The influence of stimulus size of this area was then studied in patients with several eye diseases. In many of the patients with optic neuritis, chiasmal syndrome and optic tract diseases, the stimulus size 1 was more sensitive in detecting pericentral scotoma and hemianoptic changes than the standard stimulus size 3. Thus the present results suggested that a small static target such as the stimulus size 1 should be used to examine the pericentral visual field abnormalities in neuro-ophthalmological cases.
- 1989-03-25
論文 | ランダム
- 国際 第33回UJNR海底調査専門部会に出席して
- 中国のベトナム侵略と社会主義の大義
- 高速鉄道車両用セミアクティブサスペンション開発小史
- Fourier Analysis of Interference Structure in X-Ray Specular Reflection from Thin Films
- 大分県の四つがな弁略報--玖珠町木牟田方言について