<原報>アンモニアガス電極とカリウムイオン電極を用いるアンモニウムイオンおよびカリウムイオンの分別定量
スポンサーリンク
概要
- 論文の詳細を見る
Ammonium ion in the mixture can be determined by the measurement of electromotive force of Orion ammonia gas-sensing electrode, after the ion is changed with 0.1M NaOH to ammonia in the solution. Potassium ion in the mixture can be also determined in 0.01M Tris-HCl buffer, pH 7.5,can be also measured Orion potassium ion indicator and Orion double-junction reference electrodes. The response of the both indicator electrodes are linear over the concentration ranges from (10)^<-1>∿3×(10)^<-6>M and a Nernstian slope of 58mV per concentration decade at 25.0℃. (Tables I and IV, Figs. 2 and 5) The influence of the coexisting ions are represented in Tables II, III and V and Figs. 3∿9. Ten times of K^+ in the sample do not interfere the determination of (NH_4)^+ with the ammonia gas-sensing electrode, and also Ca(II) or Mg(II) in the solution can be masked with EDTA. Sodium and lithium ions do not quite disturb the measurement of K^+ with the potassium ion electrode, and also equimolar concentration of Ca(II) or Mg(II) in the sample solution do not interfere.