Spectrophotometer Calibration by a Double Integrating Sphere Reference Light Source and Display Panel Measurement Using Dark Sphere
スポンサーリンク
概要
- 論文の詳細を見る
We succeeded to develop a reference light source in the range of very low luminance using a double integrating sphere system, and calibrated a commercial spectrophotometer below 1 × 10-5cd/m2 levels, which is 1/100 lower than the specified limit for measurement. And we improved measurements in the ultra low luminance range of displays using the calibrated commercial spectrophotometer and a dark sphere to suppress the influence of the surround.
- 2010-11-01
著者
-
Shimura Tsutomu
The Univ. Of Tokyo
-
Kubota Shigeo
The Univ. Of Tokyo
-
MATSUMOTO Tatsuhiko
Sony
-
HAGA Shuichi
Sony
-
NAKATSUE Takehiro
Sony
-
OHSAKO Junichi
The Univ. of Tokyo