A Comparison of Mammographic X-ray Spectra : Simulation with EGSnrc and Experiment with CdTe Detector
スポンサーリンク
概要
- 論文の詳細を見る
Mammographic x-ray spectra simulated by BEAMnrc/EGSnrc Monte Carlo code were qualitatively compared with the results obtained from the direct measurement using a cadmium telluride x-ray spectroscopy system and from the generation of IPEM report number 78. Generally, there is good agreement between the simulated and measured spectra, though there are slight differences at low energy in which the K-characteristic x-ray intensity is relatively higher for IPEM spectra. In addition, transmission curves were measured and simulated using a breast tissue-equivalent phantom (BR-12) as filtration. Comparison of the transmission curves shows good agreement. Moreover, the first half value layer (HVL) from direct measurement using ion chamber was consistent with the first HVL calculated by simulated spectra. Therefore, Monte Carlo may be used as an alternative tool for obtaining x-ray spectra when direct measurement is not available.
- Journal of Radiation Research 編集委員会の論文
- 2009-11-16
著者
-
NIGAPRUKE Khatayut
Department of Physics, Faculty of Science, Khon Kaen University
-
PUWANICH Patana
Department of Physics, Faculty of Science, Khon Kaen University
-
PHAISANGITTISAKUL Nakorn
Department of Physics, Faculty of Science, Chulalongkorn University
-
YOUNGDEE Wiwat
Department of Physics, Faculty of Science, Khon Kaen University
-
Youngdee Wiwat
Department Of Physics Faculty Of Science Khon Kaen University
-
Puwanich Patana
Department Of Physics Faculty Of Science Khon Kaen University
-
Nigapruke Khatayut
Department Of Physics Faculty Of Science Khon Kaen University
-
Phaisangittisakul Nakorn
Department Of Physics Faculty Of Science Chulalongkorn University
関連論文
- Monte Carlo Simulation of Average Glandular Dose and an Investigation of Influencing Factors
- A Comparison of Mammographic X-ray Spectra : Simulation with EGSnrc and Experiment with CdTe Detector