球形光束計における光源ユニット自己吸収の分光補正の検討
スポンサーリンク
概要
- 論文の詳細を見る
An integrating sphere is used to measure spectral distribution, calculate total luminous flux and calculate chromaticity coordinates of LEDs. It usually has to consider self-absorption when light source is radiating in the integrating sphere. The LED with coloured parts, such as printed circuit board, heat sink and so on, should be careful in particular. In case of the large enough sphere, self-absorption spectrum of LED has almost flat spectral distribution and its absorption level is few. Hence, the calculated total luminous flux and chromaticity coordinates of LED are almost same, even if self-absorption correction is applied or not. In the small sphere case, self-absorption of LED has spectral distribution with coloured piece and is given some influence. The correction spectrum can be measured individually in each color peace with an auxiliary lamp set up in the integrating sphere. After having spectral correction, total luminous flux comes close to large sphere case and chromaticity coordinates accord with large sphere case.
- 社団法人 電気学会の論文
- 2008-11-01
著者
関連論文
- 積分半球を用いた光源の全光束測定
- 球形光束計における光源ユニット自己吸収の分光補正の検討
- 2B12 LS-LCDの光散乱特性評価装置の開発
- 2B11 LS-LCDの照射角度依存性の評価
- 2B10 LS-LCDの光散乱特性のシミュレーション
- 2B09 LS-LCDのレーザー光散乱法によるミクロ構造の評価
- 3D124 光散乱型液晶の測定と評価
- 175.ポリクロメータによる光源色の測定((9)測光・測色・放射測定)
- LED-16 全光束モードにおけるLEDスペクトル測定に対する補正方法
- 球形光束計用受光器の斜入射光特性の検討
- 分光測定・放射測定および光応用計測
- 5.5 分光測定・放射測定および光応用計測(第5章 光・放射の計測)
- 球形光束計用受光器の斜入射光特性の検討
- 分光測定・放射測定および光応用計測
- 5.5 分光測定・放射測定および光応用計測(第5章 光放射の計測)