ツェナー標準電圧発生器の出力端子と筐体間に発生する絶縁抵抗の湿度特性の評価
スポンサーリンク
概要
- 論文の詳細を見る
Four Zener voltage standards ( Fluke 732B ) belonging to National Metrology Institute of Japan (NMIJ), AIST have contained in a particularly designed box that is called an “up-graded box". This box need for enhancing the insulation resistance occurred between the 1.018 V output terminal and chassis of the Zener voltage standard, because the 1.018 V output voltages have 1kΩ output-resistance. Hence, affect refer to best measurement capability 4 nV (k=1) of the 1.018 V output voltage arises from load effect between the insulation resistance and 1kΩ output-resistance. Moreover, this insulation resistance has an effect on the environmental humidity-change of 40%RH, 50%RH, 60%RH, 65%RH and 70%RH. Therefore, this paper revealed type B standard uncertainty of the 1.018 V output voltages arose by the humidity characteristic of the insulation resistance.
- 社団法人 電気学会の論文
- 2006-12-01