A Universal X-ray Generator for Biomedical Radiography (II):Generation of Repetitional Pulsed X-rays
スポンサーリンク
概要
- 論文の詳細を見る
A method for generating repetitional pulsed x-rays using a universal x-ray generator having a cold cathode radiation tube is described. This generator consisted of the following major components: a high speed power supply, a repetitional high voltage pulse generator, a turbo molecular pump, and an oil-cooled x-ray tube. The pulse height became higher according to increases in the condenser charging voltages, and the widths were about 5Ons. The repetitional rate was primarily determined by the current capacity of the power supply and the time constant (τ=CR) of the pulser circuit, and its value was than 20Hz.<BR>The x-ray quality primarily became hard by increasing the charging voltage and by inserting metal filters. The effective focal spot size primarily varied according to the diameter of the anode tip, and its size was less than 3.0mm in diameter. Pulsed x-ray fluoroscopy was performed by using an image intensifier system for medical use.
- Medical Imaging and Information Sciencesの論文
Medical Imaging and Information Sciences | 論文
- 系列依存性による2次元画像の解析およびその周辺
- 小型X線マイクロビームスキャナによるディジタル画像
- EMIスキヤン像のデータ処理-アナログメモリを用いた多重ウィンドウ表示と多重スライス表示-
- RadiographにおけるGraininessとSharpnessの検討
- X線用イメジインテンシファイアの雑音について