Variation of Wiener Spectra in Contents of Scattered Radiation
スポンサーリンク
概要
- 論文の詳細を見る
The visual perception of image signals in radiography are greatly deteriorated because scattered radiation causes a low contrast image. The contrast immediately decreases the value of from 1.0 to 0.3when the content of scattered radiation changes from 0 to 0.5.<BR>In the report, the Wiener spectra were determined by different a content of the scattered radiation controlled field size. Magnitude of the Wiener spectrum at a large field size(300×300mm<SUP>2</SUP>)is 1ower than those at small(40×40mm<SUP>2</SUP>), and slit(4×300mm<SUP>2</SUP>)field sizes up to the spatial the spatial frequency of 10 cycles/mm. X-ray absorption with a screen at a large field size is 0.361 and at a small field is 0.285.Namely, alot of X-ray absorption at a large field size leads a low quantum mottle.