Evaluation of Signal Detectability with Lung Nodular Reproduction by ROC Analysis using Photographic Density
スポンサーリンク
概要
- 論文の詳細を見る
We examined the effect that contrast and graininess of X - ray film/screen system had on the detection of the faint nodular shadows which overlaps with ribs and mediastinum. To evaluate the system, we adopted the receiver operating characteristic (ROC) analysis by the continuously - distributed scale method which using photographic densities of the shadow parts of chest X - rays as data. As a result, the detectability of the screen/film system for chest radiography was significantly higter than that of the conventional system. The result also agreed well with number of noise equivalent quata (NEQ(u)). Therefore, we conclude that this evaluation method is more convenient than NEQ(u), and it seems to be an useful evaluation method integrating contrast and graininess.
- Medical Imaging and Information Sciencesの論文
Medical Imaging and Information Sciences | 論文
- 系列依存性による2次元画像の解析およびその周辺
- 小型X線マイクロビームスキャナによるディジタル画像
- EMIスキヤン像のデータ処理-アナログメモリを用いた多重ウィンドウ表示と多重スライス表示-
- RadiographにおけるGraininessとSharpnessの検討