Reliability of Ultrasound Duplex for Detection of Hemodynamically Significant Stenosis in Hemodialysis Access
スポンサーリンク
概要
- 論文の詳細を見る
Objective: This study aims to evaluate the accuracy of AVF and AVG duplex ultrasound (US) compared to angiographic findings in patients with suspected failing dialysis access.Materials and Methods: From July 2008 to December 2010, US was performed on 35 hemodialysis patients with 51 vascular accesses having clinical feature or dialysis parameter suspicious of access problem. Peak systolic velocity ratio of ≥2 was the criteria for diagnosing stenosis ≥50%. Fistulogram was performed in all these patients. Results of US and fistulogram were compared using Kappa and Receiver Operator Characteristic (ROC) analyses.Results: In 51 accesses (35 AVF, 16 AVG), US diagnosed significant stenosis in 45 accesses according to the criteria and angiogram confirmed 44 significant stenoses. In AVF lesions, Kappa was 0.533 with 93.3% sensitivity and 60% specificity for US whereas in AVG lesions, Kappa was 0.636 with 100% sensitivity and50% specificity. Overall Kappa value of 0.56 meant fair to good agreement. ROC demonstrated area under the curve being 0.79 for all cases and was significant (p= 0.016). Using the ≥50% criteria for stenosis diagnosed by US yielded the best sensitivity (95.5%) and specificity (57.1%).Conclusion: Duplex ultrasound study, using ≥50% criteria, is a sensitive tool for stenosis detection in patients with suspected failing AVF and AVG.
- 日本脈管学会、日本血管外科学会、日本静脈学会の論文
日本脈管学会、日本血管外科学会、日本静脈学会 | 論文
- Multidetector-row CT Angiography of Lower Extremities: Usefulness in the Diagnosis of and Intervention for Peripheral Arterial Disease
- Inflammatory Abdominal Aortic Aneurysm: Report of Seven Cases
- Effects of Antiplatelet Agents and Other Factors on Neointimal Proliferation in Iliac Artery Stenting: Intravascular Ultrasound Analysis
- A Case Report of Aneuysmectomy after Thrombo-Intimectomy for Spontaneous Isolated Superior Mesenteric Artery Dissection
- Evaluation of Fibrous Cap Rupture of Atherosclerotic Carotid Plaque with Thin-Slice Source Images of Time-of-Flight MR Angiography