Diagnostic MR imagings of temporomandibular joint disorders by 1.5 tesla superconducting magnet.
スポンサーリンク
概要
- 論文の詳細を見る
For diagnostic imaging of temporomandibular joint (TMJ) disorders, 1.5 Tesla superconducting magnet system was applied in 65 patients with 75 joints. MR imaging was performed with GE-Signa 1.5 Tesla MR imaging system with the body coil used as the transmitter and 3 inch surface coil as the receiver. For detecting T<SUB>1</SUB> weighted imaging, the following parameters were used; TR: 1000, TE: 25, Excitation: 1. Data matrix: 256×128, FOV: 8 cm, Slice thickness: 3 mm. For taking T<SUB>2</SUB> weighted imaging, TR: 2200, TE: 80 was used.<BR>For anatomical interpretation of TMJ, 3-mm thick images with a 1-mm interspace sagittal multiplane with TR: 1000/TE: 25 sequence was useful, and additional coronal images were beneficial for examining condylar morphology as well as mediolateral disk position. T<SUB>2</SUB>weighted imaging obtained the joint effusion suspected of intra-articular inflammation. These findings suggest that MR imaging is useful in the detection and evaluation of TMJ patho-morphology.