A prototype of automated blood cell analyzer has been developed. The process is performed by the pattern recognition method.<BR>Four types of examinations are automatically operated : <BR>1) differential count of normal and abnormal white blood cells, <BR>2) morphological analysis of red blood cells, <BR>3) numerical platelet estimates and<BR>4) reticulocyte count (a slide preparation differing from other three).<BR>The WBC differential count is the essential of the process and applies an adaptive and hierarchical classifier.fw Here, the word "adaptive" means self compensation using standard cells in each sample. The standard cells are automatically picked up from input cells.<BR>The WBC differential count is processed at the rate of one sample per minute. Net discriminating rate of normal WBC differential count is better than 98%.