Development of a New Method for Calibration of Pneumotachograph According to Flow Rate. Evaluation of a computer-controlled-system for assessing respiratory gas exchange using this method.
スポンサーリンク
概要
- 論文の詳細を見る
1) We evaluated the method for calibrating the pneumotachograph according to flow rate by the software on personal computer. Flow volume was detected using the approximated equation (y = 1.294x<SUP>2</SUP>+0.989x+35.824, r = 0.976, p < 0.01) from various mean flow rate and integration of output voltage. The error between syringe volume (21) and the estimated value on the equation was l0d±8 ml (0.48±0.42 %). 2 ) Using the method descrived in l), we developed computer-controlled-system for assesing respiratory gas exchange by mixing chamber method.To evaluate the accuracy of the system as compared with the Douglas Bag method, we measured the gas parameters of the both methods for two male subjects at rest(3 and 5min) and steady state exercise (3 min., 60-240watt). The error between the both methods were within±5 % in ventilation(VE), mixed expired O<SUB>2</SUB> (F<SUB>E</SUB>O<SUB>2</SUB>) and CO<SUB>2</SUB> (F<SUB>E</SUB>CO<SUB>2</SUB>) concentration, and were within±8% in O<SUB>2</SUB> uptake(VO<SUB>2</SUB>) and CO<SUB>2</SUB> production(VCO<SUB>2</SUB>). There were high positive correlation(r > 0.99, p < 0.01) between the both methods in the all parameters. These results indicate that the system in this stydy would be practical use, that VE can be measured in only one pneumotachograph, even if the range of flow rate was wide(e.g. from rest to heavy exercise).
- 日本生理人類学会の論文
日本生理人類学会 | 論文
- 呼吸同調現象(LRC)の気道閉塞圧(P_)からの検討
- 周波数変調音の擬音語表現とサイン音としての機能イメージ
- P-18 異なるコンディションでのKVA動体視力と打撃運動
- 足圧中心動揺と筋力の関係から見た高齢者の側方ステップ反応特性
- 高齢者における重心側方移動時の前額面上姿勢方略特性について