Indoor <SUP>222</SUP>Rn Concentrations in Radioisotope Use, Uranium Use, and General Use Laboratories
スポンサーリンク
概要
- 論文の詳細を見る
Indoor <SUP>222</SUP>Rn concentration was measured using an electrostatic radon monitor of PMT-TEL (Pylon Co.). The variation of radon concentration was compared among three kinds of laboratories: Radioisotope (RI) use, Uranium-use and general use laboratories.<BR>The <SUP>222</SUP>Rn concentration observed in RI Center under air conditioning was about 3 Bq/m<SUP>3</SUP> close to the outside <SUP>222</SUP>Rnlevel. The indoor radon level increased up to about 30 Bq/m<SUP>3</SUP>after the air conditioning was off. Radon exhalation rate was evaluated from increasing rate of radon concentrations after the air conditioning was stopped.<BR>In uranium use laboratories, the radon concentration was 40-70 Bq/m<SUP>3</SUP> due to the storage of uranium compounds (about 12 kg). The negative correlation between radon concentration and atmospheric pressure was obtained when the variations of indoor temperature and relative humidity were small. On the other hand, the radon concentration was 5-15 Bq/m<SUP>3</SUP> in general use laboratories of old 2nd building of Engineering, which had not so well shielded windows.<BR>In conclusion, indoor radon concentration in uranium use laboratories and RI center became high after air conditioning was off, and so it is recommended that the ventilation should be operated before entering the uranium use laboratories and RI center with well shielded facilities.
- 日本放射線安全管理学会の論文
日本放射線安全管理学会 | 論文
- イメージングプレートに記録されたPSL強度分布の解析
- イメージングプレートにおけるPSL強度分布に着目した線種識別法の提案
- 線量当量(Dose Equivalent)と等価線量(Equivalent Dose)
- 広帯域エネルギー対応マルチ放射線モニタDARWINの開発 (第4回JPSMシンポジウム) -- (先端放射線計測技術と放射線安全管理--測定の自動化と広帯域化)
- 半導体式電子ポケット線量計の経年変化と方向特性