Determination of selenium (IV) in water by graphite furnace atomic absorption spectrometry after preconcentration with microorganism.
スポンサーリンク
概要
- 論文の詳細を見る
The determination method of selenium (IV) in water samples by graphite furnace atomic absorption spectrometry (GF-AAS) after pre-concentration with bacteria was developed. High Se (IV) -absorbing ability was found in <I>Bacillus</I> sp. which was separated from specific microbes for wastewater treatment systems. Precipitated bacteria was dissolved with 0.7ml of HNO<SUB>3</SUB> (30%) and 0.3ml of palladium solution (1000mg/l) . The dissolved solution was injected into the graphite tube and Se peak height was measured under the optimum GF-AAS operating conditions. No interfering effects were found for coexisting 100μg/l of various metal ions on the l0μg/l of Se (IV) analysis. On the other hand, the depressant effects were caused by 1000μg/l of metal anions. By the proposed method, 0.5-30μg/l of Se (IV) could be quantitatively recovered and determined with the relative standard deviation of 2.2%. This method was applied to the determination of Se (IV) in wastewaters and the results were in good agreement with them by the Cu-DDTC coprecipitation/GF-AAS method. Pre-concentration of Se (IV) from gas washer wastes was not possible with this method due to the high salinity of the solutions.
- 環境技術学会の論文
環境技術学会 | 論文
- 消化液農地投与による環境保全
- 畜産廃水の土壌浸透システムにおける飼料作物の育成 : 土壌中のリン・栄養塩類の挙動および塩類集積に関する考察
- (株)ユニチカ環境技術センター見学会
- 水環境中での砒素の挙動・特性に対応した除去方法の選択 : 吸着法を中心に
- 高温高圧条件下における2,4-ジクロロフェノキシ酢酸の分解