Breakage of ABS Cup in Ceramic-on-Ceramic Total Hip Arthroplasty: a case report
スポンサーリンク
概要
- 論文の詳細を見る
We describe a case of cup breakage in a ceramic-on-ceramic total hip arthroplasty at 3 years postoperatively. In 1999, a 68-year-old woman recieved a left cementless total hip arthroplasty for osteoarthritis, using a ceramic-on-ceramic prosthesis (Kyocera ABS cup, alumina ball) . In May, 2002, she noted a left groin creak which progressively increased. A radiograph of the pelvis revealed displacement of the cup liner, and a revisional operation was performed in August, 2002. Analysis of retrieved implants suggested that increased friction on the alumina bearing surface, possibly due to undetermined interposition, resulted in increased torque, and subsequently the liner was sheared off the metal shell. Since there have been increasing ABS cup failures, careful follow-up of operated patients is needed for early detection of breakage.
- 日本関節病学会の論文
日本関節病学会 | 論文
- 関節リウマチ患者における手術と機能回復
- 低出力超音波刺激(LIPUS)および線維芽細胞増殖因子(FGF-2)の骨形成促進作用におけるヒアルロン酸(HA)の影響
- 変形性関節症に対するグルコサミン・コンドロイチンの有用性 (第36回[日本関節病]学会寄稿)
- 人工膝関節再置換術の経験--脛骨結節骨切り術の適応と合併症について
- RA頚椎病変に対する上位頚椎後方除圧固定術の治療成績と問題点