On High Vacuum Standard (2nd Report):Factors Influencing the Calibration of Sub-Standard Ionization Gauges
スポンサーリンク
概要
- 論文の詳細を見る
In order to determine the sub-standard ionization gauges, many factors influencing the sensitivity of the gauges were investigated in detail.<BR>For the calibration of the sub-standard gauges with the standard McLeod gauge, it is required that the deviations of the sensitivity is within one per cent in the pressure range of 10<SUP>-4</SUP> to about 10<SUP>-3</SUP> <I>mm</I>. of mercury. <BR>Experiments have been carried out under many conditions.<I>i. e.</I> changing the values of pressure, sorts of gases, electron currents and potentials of electrodes, for several types and dimensions of gauges. In these experiments, we obtained an optimum condition for a gauge having sensitivity constancy for pressures up to 3×10<SUP>-3</SUP> mm of mercury while we found such an unfavourable phenomena that the stability tended to decrease slowly during five to twenty-five hours after the outgassing procedure of electron bombardment.