Retrograde Cerebral Perfusion Using a New Double-Lumen Balloon Catheter via Internal Jugular Vein Cannulation.
スポンサーリンク
概要
- 論文の詳細を見る
We developed a new double-lumen balloon catheter for retrograde cerebral perfusion (RCP) via jugular vein cannulation. Between November 1996 and September 2000, 34 of 73 patients treated with surgical procedures for thoracic aortic aneurysms underwent RCP using the new catheter during circulatory arrest under deep hypothermia. Nine patients underwent a median sternotomy, and 25 underwent a left thoracotomy. In all cases, the new catheter installation under fluoroscopy was easy, and it took about 15min. The mean RCP time, pressure, and flow rate were 26.8min, 20.0mmHg, and 202.6ml/min, respectively. Our procedure using the new catheter was safe and easy in RCP during circulatory arrest in aortic arch replacement regardless of surgical approaches such as a left thoracotomy or median sternotomy.
- 特定非営利活動法人 日本心臓血管外科学会の論文
特定非営利活動法人 日本心臓血管外科学会 | 論文
- 慢性静脈不全症に対する深部静脈逆流遮断手術 弁形成術と弁置換術
- 高令者(70才以上)の腹部大動脈外科の検討
- 大腿吻合部動脈瘤の発生原因の検討
- High riskの潜在性凝固異常と腹部大動脈瘤手術
- Aorto-femoro-poplitealの広汎閉塞に対する2期的再建法の工夫