ガンマカメラ用57Co面線源の基礎的検討と日常点検への利用
スポンサーリンク
概要
- 論文の詳細を見る
Quality control (QC) detects changes in the performance of gamma cameras that could adversely affect interpretations of clinical studies. We used plate and sheet 57Co flood sources to measure extrinsic uniformity during daily QC. Each source, when placed on the top of a collimated detector, allowed the acquisition of uniform images from both detectors, thus reducing the amount of time needed to perform daily QC. No serious problems with the gamma camera system were revealed by visual checks, and changes in detector sensitivity were rapidly determined by observing daily variations in the measured values of extrinsic uniformity. Furthermore, 57Co flood sources confer advantages in that they shorten the time required for preparation of flood sources and reduce the consequent exposure of medical staff to radiation.