A Study of Availability of Paraffin Wax for Valve in Microfluidic Channel
スポンサーリンク
概要
- 論文の詳細を見る
Availability of a novel micro-valve made of paraffin wax which is used in a "swallowable biosensor" is studied. The valve was composed of a plug made of paraffin wax which sealed a micro fluid channel and micro heater to melt the paraffin plug. Additionally, a collecting area of paraffin after melting was formed downstream of the paraffin valve to prevents the melted paraffin from influencing on another devices in the swallowable biosensor. Shape of the collecting area for the melted paraffin was optimized by simulation using finite element method (FEM). As the result of simulation, it is found that two sloped weir, which is staggered each other, collected the most melted paraffin. Fabrication of the paraffin valve and the weirs collecting for the melted was also carried out. As a result of collecting test of melted paraffin, it was observed that two sloped weir collected most melted paraffin, which is the same result as the simulation.
著者
-
村上 裕二
豊橋技術科学大学 電気・電子情報工学系
-
田口 亮
関西大学 システム理工学部
-
青柳 誠司
関西大学 システム理工学部
-
村上 裕二
豊橋技術科学大学 電気・電子情報工学系
-
鈴木 昌人
関西大学 システム理工学部
-
高橋 智一
関西大学 システム理工学部
関連論文
- Blood Glucose Sensors in Digital Health-Era
- A Study of Availability of Paraffin Wax for Valve in Microfluidic Channel
- デジタルヘルスケア時代の血糖値センサ