Using a small-type crucible, the inter-laboratory comparison of the realization of the Co-C eutectic point was executed in the industrial field in Japan, with the aim to verify the possibility of thermocouple calibration at the eutectic point with an existing furnace of not so good temperature distribution: the palladium-point wire-bridge method furnace for thermocouple calibration. A small-type crucible for thermocouples was made by the National Metrology Institute of Japan, National Institute of Advanced Industrial Science and Technology (NMIJ/AIST). The inter-laboratory comparison was piloted by the Tokyo Metropolitan Industrial Technology Research Institute (TIRI), a public test and research organization in Tokyo, and was conducted among five calibration laboratories including the pilot. One of the five laboratories is a designated calibration laboratory in Japan, and the three remainders are private laboratories. Four laboratories including TIRI used existing palladium-point furnace, and one laboratory made a new furnace for this inter-laboratory comparison.
著者
-
佐藤 弘康
日本電気計器検定所
-
山嵜 春樹
田中貴金属工業
-
佐々木 正史
(地独) 東京都立産業技術研究センター
-
木村 和弘
(株) 岡崎製作所
-
山田 善郎
(独) 産業技術総合研究所
-
沼尻 治彦
(地独) 東京都立産業技術研究センター
-
相川 孝生
(株) チノー
-
石井 順太郎
(独) 産業技術総合研究所
関連論文
- 既存炉を利用した白金パラジウム熱電対の不均質測定に関する研究
- 小型簡易定点炉によるCo-C共晶点の実現
- GTH(ごっつう・つよい・はっきん)の高温特性 (特集 白金族金属の資源と製造・リサイクル技術の新展開)
- るつぼ法とワイヤ法によるパラジウム定点の信頼性評価
- R熱電対の高温特性
- 熱電対不均質測定装置の評価
- ワイヤーブリッジ法のプラトー形状改善のための検討
- Pd点溶融法におけるR熱電対のドリフトと不均質の調査
- 校正の繰り返しによる熱電対のドリフトと不均質の調査
- 日本の温度トレーサビリティの現状と今後の動向
- 工業用白金抵抗温度計の1000時間安定性試験
- 熱電対用小型共晶点セルの持回り試験
- Inter-laboratory Comparison of the Co-C Eutectic Point with Small-type Cell for Thermocouples
- 熱電対用小型共晶点セルの持回り試験
- 工業用白金抵抗温度計の1000時間安定性試験
- R熱電対の不均質評価方法の検討
- 熱電対用小型共晶点セルの持回り試験
- 酸化物分散強化白金熱電対(TEMPLAT^)の開発